DICP OpenIR
A vacuum assisted dynamic evaporation interface for two-dimensional normal phase/reverse phase liquid chromatography
Ding, Kun; Xu, Yuan; Wang, Hua; Duan, Chunfeng; Guan, Yafeng; Guan YF(关亚风)
关键词Solvent Evaporation Interface Loop-type Interface Two-dimensional Liquid Chromatography Band Compression Vacuum
刊名JOURNAL OF CHROMATOGRAPHY A
2010-08-20
DOI10.1016/j.chroma.2010.06.053
1217期:34页:5477-5483
收录类别SCI
文章类型Article
部门归属1
项目归属105
产权排名1;1
WOS标题词Science & Technology ; Life Sciences & Biomedicine ; Physical Sciences
类目[WOS]Biochemical Research Methods ; Chemistry, Analytical
研究领域[WOS]Biochemistry & Molecular Biology ; Chemistry
关键词[WOS]COMPLEX-MIXTURES ; SEPARATIONS
英文摘要A vacuum assisted dynamic solvent evaporation interface for coupling of two-dimensional normal phase/reverse phase liquid chromatography was developed and evaluated. A normal-phase liquid chromatographic (NPLC) column of a 250 mm x 4.6 mm I.D. 5 mu m CN phase was used as the first dimension, and a reversed-phase liquid chromatographic (RPLC) column of 250 mm x 4.6 mm I.D. 5 mu m C(18) phase was used as the second dimension. The eluent from the first dimension flowed into a fraction loop, and the solvent in the eluent was dynamically evaporated and removed by vacuum as it was entering the fraction loop of the interface. The non-evaporable analytes was retained and enriched in about 5-25 mu L solution within the loop. Up to 1 mL/min of mobile phase from the first dimension can be evaporated and removed dynamically by the interface. The mobile phase from the second dimension then entered the loop, and dissolved the concentrated analytes retained inside the loop, and carried them onto the second dimension column for further separation. The operation conditions of the two dimensions were independent from each other, and both dimensions were operated at their optimal chromatographic conditions. We evaluated the interface by controlling the loop temperature in a water bath at normal temperature, and investigated the sample losses by using standard samples with different boiling points. It was found that the sample loss due to evaporation in the interface was negligible for non-volatile samples or for components with boiling point above 340 degrees C. The interface realizes fast solvent removal of mL volume of fraction and concentration of the fraction into tenth of mu L volume, and injection of the concentrated fraction on the secondary column. The chromatographic performance of the two-dimensional LC system was enhanced without compromise of separation efficiency and selectivity on each dimension. (C) 2010 Elsevier B.V. All rights reserved.
语种英语
原文出处查看原文
WOS记录号WOS:000281291000004
引用统计
被引频次:19[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://cas-ir.dicp.ac.cn/handle/321008/103411
专题中国科学院大连化学物理研究所
通讯作者Guan YF(关亚风)
作者单位Chinese Acad Sci, Dalian Inst Chem Phys, Key Lab Separat Sci Analyt Chem CAS, Dept Instrumentat & Analyt Chem, Dalian 116023, Peoples R China
推荐引用方式
GB/T 7714
Ding, Kun,Xu, Yuan,Wang, Hua,et al. A vacuum assisted dynamic evaporation interface for two-dimensional normal phase/reverse phase liquid chromatography[J]. JOURNAL OF CHROMATOGRAPHY A,2010,1217(34):5477-5483.
APA Ding, Kun,Xu, Yuan,Wang, Hua,Duan, Chunfeng,Guan, Yafeng,&关亚风.(2010).A vacuum assisted dynamic evaporation interface for two-dimensional normal phase/reverse phase liquid chromatography.JOURNAL OF CHROMATOGRAPHY A,1217(34),5477-5483.
MLA Ding, Kun,et al."A vacuum assisted dynamic evaporation interface for two-dimensional normal phase/reverse phase liquid chromatography".JOURNAL OF CHROMATOGRAPHY A 1217.34(2010):5477-5483.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Ding, Kun]的文章
[Xu, Yuan]的文章
[Wang, Hua]的文章
百度学术
百度学术中相似的文章
[Ding, Kun]的文章
[Xu, Yuan]的文章
[Wang, Hua]的文章
必应学术
必应学术中相似的文章
[Ding, Kun]的文章
[Xu, Yuan]的文章
[Wang, Hua]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。