Study of the surface ionization detector for gas chromatography
Li, Weiwei; Wu, Dapeng; Chen, Shiheng; Peng, Hong; Guan, Yafeng; Guan YF(关亚风)
KeywordSurface Ionization Detector Gc Quartz Liner Thermo Ionization Amines Selectivity
Source PublicationJOURNAL OF CHROMATOGRAPHY A
2011-09-23
ISSN待补充
DOI10.1016/j.chroma.2011.07.074
Volume1218Issue:38Pages:6812-6816
Indexed BySCI
SubtypeArticle
Department1
Funding Project105
Contribution Rank1,1
WOS HeadingsScience & Technology ; Life Sciences & Biomedicine ; Physical Sciences
WOS SubjectBiochemical Research Methods ; Chemistry, Analytical
WOS Research AreaBiochemistry & Molecular Biology ; Chemistry
WOS KeywordORGANIC MASS-SPECTROMETRY ; AMINES
AbstractStudy of the surface ionization detector for gas chromatography; The structure of the surface ionization detector (SID) and the operation parameters of GC-SID were investigated to reduce peak tailing and to enhance sensitivity. The performances of the GC-SID, including its repeatability, linearity, sensitivity, selectivity, and tolerance towards water vapor, were evaluated systematically. Compared with nitrogen-phosphorus detector (NPD), the SID was able to detect fg level triethylamine, and selectively respond to alkylamines, some anilines, and some nitrogen heterocyclic compounds. Among alkylamines, the SID sensitivity to diisobutylamine was rather small. Even so, it was also still 10 times higher than that on NPD. The SID selectivity, defined as the sensitivity ratio between triethylamine and various tested non-nitrogen compounds, was higher than 10(6). It was found that the SID is highly tolerant towards water vapor, allowing direct injection of water sample. Finally, the GC-SID was applied to directly measure trace amines in headspace gases of rotted meat and trace simazine in tap water. The SID sensitivity to simazine was proven to be 5 times higher than that on flame ionization detector (FID). This study suggests that the SID is a promising GC detector. (C) 2011 Elsevier B.V. All rights reserved.
Language英语
WOS IDWOS:000295432700029
Citation statistics
Cited Times:1[WOS]   [WOS Record]     [Related Records in WOS]
Document Type期刊论文
Identifierhttp://cas-ir.dicp.ac.cn/handle/321008/115454
Collection中国科学院大连化学物理研究所
Corresponding AuthorGuan YF(关亚风)
AffiliationChinese Acad Sci, Dalian Inst Chem Phys, Key Lab Separat Sci Analyt Chem, Dept Instrumentat & Analyt Chem, Dalian 116023, Peoples R China
Recommended Citation
GB/T 7714
Li, Weiwei,Wu, Dapeng,Chen, Shiheng,et al. Study of the surface ionization detector for gas chromatography[J]. JOURNAL OF CHROMATOGRAPHY A,2011,1218(38):6812-6816.
APA Li, Weiwei,Wu, Dapeng,Chen, Shiheng,Peng, Hong,Guan, Yafeng,&关亚风.(2011).Study of the surface ionization detector for gas chromatography.JOURNAL OF CHROMATOGRAPHY A,1218(38),6812-6816.
MLA Li, Weiwei,et al."Study of the surface ionization detector for gas chromatography".JOURNAL OF CHROMATOGRAPHY A 1218.38(2011):6812-6816.
Files in This Item:
File Name/Size DocType Version Access License
2011TBf14H0aMF.pdf(493KB) 开放获取--Application Full Text
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[Li, Weiwei]'s Articles
[Wu, Dapeng]'s Articles
[Chen, Shiheng]'s Articles
Baidu academic
Similar articles in Baidu academic
[Li, Weiwei]'s Articles
[Wu, Dapeng]'s Articles
[Chen, Shiheng]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[Li, Weiwei]'s Articles
[Wu, Dapeng]'s Articles
[Chen, Shiheng]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.