Development of Unidirectional Negative Corona Discharge Ion Mobility Spectrometer for Detection of Explosives
Liu Ji-Wei1; Peng Li-Ying2,3; Huang Wei2; Wang Wei-Guo2; Jiang Dan-Dan2,3; Sun Qi1; Li Hai-Yang2
关键词Ion Mobility Spectrometry Explosives Ionization Source Negative Corona Discharge Unidirectional Mode
刊名CHINESE JOURNAL OF ANALYTICAL CHEMISTRY
2016-08-01
DOI10.11895/j.issn.0253.3820.160038
44期:8页:1185-1192
收录类别SCI
文章类型Article
WOS标题词Science & Technology ; Physical Sciences
类目[WOS]Chemistry, Analytical
关键词[WOS]IONIZATION SOURCE ; SENSITIVE DETECTION ; MASS-SPECTROMETRY ; RAPID DETECTION ; MICROEXTRACTION ; WATER ; AIR
英文摘要Ion mobility spectrometry (IMS) based explosives detectors is wieldy deployed at the check points of airport, etc, so the development of new ionization method to replace the radioactive Ni-63 ion source in IMS is highly demanded. In this study, a novel negative corona discharge was developed for ion mobility spectrometer for rapid detection of trace explosives, which was running in the unidirectional mode to efficiently remove nitrogen oxides and ozone produced from the discharge process. The diameter of target electrode was 3 mm, the distance between the needle and the target electrode was 2 mm, the discharge voltage was 2400 V and the flow rate of drift gas was 1200 mL/min. Under the optimized conditions, the dominant reactant ions was O-2(H2O)(n) and could be used to directly detect the common explosives, such as trinitrotoluene (TNT), ammonium nitrate (AN), nitroglycerin (NG), pentaerythritol tetranitrate (PETN) and hexahydro. 1,3,5-trinitro. 1,3,5-triazine (RDX). The limit of detection of TNT was 200 pg/mu L. These results indicated that negative corona discharge ionization source, with the advantanges of the high sensibility, simple structure and no radioactivity, could be used as a potential and promising ionization source for ion mobility spectrometry to detect explosives.
语种英语
WOS记录号WOS:000389035300004
引用统计
文献类型期刊论文
条目标识符http://cas-ir.dicp.ac.cn/handle/321008/150352
专题中国科学院大连化学物理研究所
作者单位1.Liaoning Normal Univ, Fac Chem & Chem Engn, Inst Chem Functionalized Mat, Dalian 116029, Peoples R China
2.Chinese Acad Sci, Dalian Inst Chem Phys, Dalian 116023, Peoples R China
3.Univ Chinese Acad Sci, Beijing 100049, Peoples R China
推荐引用方式
GB/T 7714
Liu Ji-Wei,Peng Li-Ying,Huang Wei,et al. Development of Unidirectional Negative Corona Discharge Ion Mobility Spectrometer for Detection of Explosives[J]. CHINESE JOURNAL OF ANALYTICAL CHEMISTRY,2016,44(8):1185-1192.
APA Liu Ji-Wei.,Peng Li-Ying.,Huang Wei.,Wang Wei-Guo.,Jiang Dan-Dan.,...&Li Hai-Yang.(2016).Development of Unidirectional Negative Corona Discharge Ion Mobility Spectrometer for Detection of Explosives.CHINESE JOURNAL OF ANALYTICAL CHEMISTRY,44(8),1185-1192.
MLA Liu Ji-Wei,et al."Development of Unidirectional Negative Corona Discharge Ion Mobility Spectrometer for Detection of Explosives".CHINESE JOURNAL OF ANALYTICAL CHEMISTRY 44.8(2016):1185-1192.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Liu Ji-Wei]的文章
[Peng Li-Ying]的文章
[Huang Wei]的文章
百度学术
百度学术中相似的文章
[Liu Ji-Wei]的文章
[Peng Li-Ying]的文章
[Huang Wei]的文章
必应学术
必应学术中相似的文章
[Liu Ji-Wei]的文章
[Peng Li-Ying]的文章
[Huang Wei]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。