关键词[WOS]PERFORMANCE LIQUID-CHROMATOGRAPHY ; RIGID POLYMER MONOLITHS ; SEPARATION MEDIA ; STATIONARY PHASES ; HOMOLOGOUS SERIES ; CONTINUOUS BEDS ; COLUMNS ; BEHAVIOR ; MEMBERS
英文摘要Polymethacrylate-based monolithic columns were prepared for capillary electrochromatography (CEC) by in situ copolymerization of butyl methacrylate (BMA), 2-acrylamido-2-methyl-1-propanesulfonic acid (AMPS), and ethylene dimethacrylate (EDMA) in the presence of a porogen in fused-silica capillaries of 100 mum I.D. The abnormal phenomenon that retention factors for neutral species decreases with applied voltage in CEC was observed. Capillary electrophoresis (CE) instruments usually require a period of time to increase voltage from 0 kV to desired value, which is called as ramp time. Such ramp time and any error in the determination of dead time should be taken into account during the accurate calculation of retention factors. After the correction of the retention factors, the plots of the corrected factors for alkylbenzene versus applied voltage were made, the absolute value of the plot slopes are less than 1.8 X 10(-4), Which indicates that the corrected retention times for neutral species do not show any dependence on applied voltage. Further, the plots of the corrected retention times for acidic and basic compounds versus the reciprocal of applied voltage were drawn, where the target compounds were eluted in neutral form. The very nice linearity of the plots was obtained. The linear correlation coefficients are over 0.999. Here, the slopes of the plots represent
语种英语
原文出处查看原文
WOS记录号WOS:000183802500012
引用统计
被引频次:5[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://cas-ir.dicp.ac.cn/handle/321008/83433
专题中国科学院大连化学物理研究所
作者单位1.Acad Sinica, Dalian Inst Chem Phys, Natl Chromatog Res & Anal Ctr, Dalian 116011, Peoples R China
2.GSF, Natl Ctr Environm & Hlth, Inst Ecol Chem, D-85764 Neuherberg, Germany
推荐引用方式
GB/T 7714
Ping, GC,Zhang, WB,Zhang, L,et al. Abnormal phenomenon of the dependence of the retention factors for uncharged species on applied voltage in capillary electrochromatography[J]. CHROMATOGRAPHIA,2003,57(11-12):777-781.
APA Ping, GC.,Zhang, WB.,Zhang, L.,Zhang, LH.,Schmitt-Kopplin, P.,...&Zhang, YK.(2003).Abnormal phenomenon of the dependence of the retention factors for uncharged species on applied voltage in capillary electrochromatography.CHROMATOGRAPHIA,57(11-12),777-781.
MLA Ping, GC,et al."Abnormal phenomenon of the dependence of the retention factors for uncharged species on applied voltage in capillary electrochromatography".CHROMATOGRAPHIA 57.11-12(2003):777-781.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
2003790MQ5939A.pdf(616KB) 开放获取--请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Ping, GC]的文章
[Zhang, WB]的文章
[Zhang, L]的文章
百度学术
百度学术中相似的文章
[Ping, GC]的文章
[Zhang, WB]的文章
[Zhang, L]的文章
必应学术
必应学术中相似的文章
[Ping, GC]的文章
[Zhang, WB]的文章
[Zhang, L]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。